| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.121 | 2.035 | 
| 500.00 | 1.579 | 1.493 | 
| 250.00 | 1.216 | 1.130 | 
| 125.00 | 0.782 | 0.696 | 
| 62.50 | 0.427 | 0.341 | 
| 31.25 | 0.318 | 0.232 | 
| 15.63 | 0.217 | 0.131 | 
| 0.00 | 0.086 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 78-92% | 85% | 
| EDTA plasma(n=5) | 92-106% | 99% | 
| Heparin plasma(n=5) | 87-95% | 91% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 93-106% | 87-99% | 85-93% | 79-96% | 
| EDTA plasma(n=5) | 97-105% | 84-96% | 87-98% | 95-102% | 
| Heparin plasma(n=5) | 95-103% | 79-97% | 93-102% | 87-103% |