| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.054 | 1.972 | 
| 500.00 | 1.638 | 1.556 | 
| 250.00 | 1.032 | 0.950 | 
| 125.00 | 0.733 | 0.651 | 
| 62.50 | 0.548 | 0.466 | 
| 31.25 | 0.335 | 0.253 | 
| 15.63 | 0.173 | 0.091 | 
| 0.00 | 0.082 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 82-98% | 90% | 
| EDTA plasma(n=5) | 90-107% | 98% | 
| Heparin plasma(n=5) | 87-99% | 93% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-94% | 87-96% | 82-95% | 93-102% | 
| EDTA plasma(n=5) | 95-106% | 85-104% | 90-98% | 82-98% | 
| Heparin plasma(n=5) | 88-97% | 96-102% | 81-93% | 89-97% |