| 48 T | $412.00 | 
| 96 T | $588.00 | 
| 96 T*5 | $2499.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 500.00 | 2.096 | 2.023 | 
| 250.00 | 1.497 | 1.424 | 
| 125.00 | 1.121 | 1.048 | 
| 62.50 | 0.752 | 0.679 | 
| 31.25 | 0.427 | 0.354 | 
| 15.63 | 0.305 | 0.232 | 
| 7.82 | 0.232 | 0.159 | 
| 0.00 | 0.073 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 82-94% | 88% | 
| EDTA plasma(n=5) | 97-107% | 102% | 
| Heparin plasma(n=5) | 87-99% | 93% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 95-104% | 79-92% | 82-95% | 86-97% | 
| EDTA plasma(n=5) | 86-93% | 91-105% | 87-98% | 99-105% | 
| Heparin plasma(n=5) | 88-102% | 86-92% | 81-93% | 89-104% |