| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.035 | 1.952 | 
| 5.00 | 1.602 | 1.519 | 
| 2.50 | 1.121 | 1.038 | 
| 1.25 | 0.735 | 0.652 | 
| 0.63 | 0.548 | 0.465 | 
| 0.32 | 0.315 | 0.232 | 
| 0.16 | 0.179 | 0.096 | 
| 0.00 | 0.083 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 95-105% | 100% | 
| EDTA plasma(n=5) | 80-95% | 87% | 
| Heparin plasma(n=5) | 92-105% | 98% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 83-94% | 85-96% | 92-103% | 86-97% | 
| EDTA plasma(n=5) | 83-96% | 81-94% | 86-98% | 92-101% | 
| Heparin plasma(n=5) | 83-96% | 86-95% | 90-102% | 81-93% |