| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.135 | 2.068 | 
| 5.00 | 1.472 | 1.405 | 
| 2.50 | 1.139 | 1.072 | 
| 1.25 | 0.839 | 0.772 | 
| 0.63 | 0.509 | 0.442 | 
| 0.32 | 0.319 | 0.252 | 
| 0.16 | 0.232 | 0.165 | 
| 0.00 | 0.067 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 86-98% | 92% | 
| EDTA plasma(n=5) | 95-107% | 101% | 
| Heparin plasma(n=5) | 83-95% | 89% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-94% | 79-96% | 92-101% | 83-90% | 
| EDTA plasma(n=5) | 79-92% | 95-103% | 82-98% | 85-97% | 
| Heparin plasma(n=5) | 88-97% | 86-92% | 81-93% | 89-105% |