| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 50000.00 | 0.175 | |
| 25000.00 | 0.367 | |
| 12500.00 | 0.577 | |
| 6250.00 | 0.833 | |
| 3125.00 | 1.197 | |
| 1562.50 | 1.578 | |
| 781.25 | 1.755 | |
| 0.00 | 2.101 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 92-106% | 99% | 
| EDTA plasma(n=5) | 88-103% | 95% | 
| Heparin plasma(n=5) | 80-97% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 95-103% | 79-97% | 96-105% | 95-103% | 
| EDTA plasma(n=5) | 85-92% | 79-96% | 95-103% | 87-98% | 
| Heparin plasma(n=5) | 85-92% | 89-101% | 87-98% | 85-92% |