| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 400.00 | 1.994 | 1.905 | 
| 200.00 | 1.565 | 1.476 | 
| 100.00 | 0.957 | 0.868 | 
| 50.00 | 0.505 | 0.416 | 
| 25.00 | 0.433 | 0.344 | 
| 12.50 | 0.286 | 0.197 | 
| 6.25 | 0.251 | 0.162 | 
| 0.00 | 0.089 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 78-95% | 86% | 
| EDTA plasma(n=5) | 89-103% | 96% | 
| Heparin plasma(n=5) | 86-99% | 92% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-104% | 97-106% | 86-97% | 92-99% | 
| EDTA plasma(n=5) | 85-93% | 98-105% | 87-98% | 95-102% | 
| Heparin plasma(n=5) | 98-102% | 86-92% | 81-93% | 89-98% |