| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 50.00 | 2.336 | 2.228 | 
| 25.00 | 1.566 | 1.458 | 
| 12.50 | 0.913 | 0.805 | 
| 6.25 | 0.739 | 0.631 | 
| 3.13 | 0.385 | 0.277 | 
| 1.57 | 0.305 | 0.197 | 
| 0.79 | 0.226 | 0.118 | 
| 0.00 | 0.108 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 83-95% | 89% | 
| EDTA plasma(n=5) | 80-96% | 88% | 
| Heparin plasma(n=5) | 80-95% | 87% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 95-104% | 97-105% | 82-95% | 89-97% | 
| EDTA plasma(n=5) | 82-95% | 89-103% | 79-91% | 85-93% | 
| Heparin plasma(n=5) | 78-91% | 86-92% | 89-96% | 91-97% |