 
| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 8000.00 | 1.983 | 1.897 | 
| 4000.00 | 1.628 | 1.542 | 
| 2000.00 | 1.318 | 1.232 | 
| 1000.00 | 0.914 | 0.828 | 
| 500.00 | 0.548 | 0.462 | 
| 250.00 | 0.357 | 0.271 | 
| 125.00 | 0.215 | 0.129 | 
| 0.00 | 0.086 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 80-92% | 86% | 
| EDTA plasma(n=5) | 78-92% | 85% | 
| Heparin plasma(n=5) | 87-99% | 93% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-99% | 82-96% | 92-101% | 82-90% | 
| EDTA plasma(n=5) | 88-96% | 90-101% | 93-102% | 85-94% | 
| Heparin plasma(n=5) | 89-97% | 86-95% | 81-93% | 91-103% |