| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 20000.00 | 0.191 | |
| 10000.00 | 0.396 | |
| 5000.00 | 0.621 | |
| 2500.00 | 0.861 | |
| 1250.00 | 1.119 | |
| 625.00 | 1.522 | |
| 312.50 | 1.799 | |
| 0.00 | 2.132 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 83-99% | 95% | 
| EDTA plasma(n=5) | 85-97% | 91% | 
| Heparin plasma(n=5) | 97-105% | 101% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 80-101% | 87-96% | 86-97% | 88-95% | 
| EDTA plasma(n=5) | 85-92% | 79-96% | 92-101% | 87-98% | 
| Heparin plasma(n=5) | 95-103% | 89-96% | 87-98% | 85-93% |