| 48 T | $412.00 | 
| 96 T | $588.00 | 
| 96 T*5 | $2499.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 2000.00 | 1.952 | 1.866 | 
| 1000.00 | 1.754 | 1.668 | 
| 500.00 | 1.484 | 1.398 | 
| 250.00 | 0.795 | 0.709 | 
| 125.00 | 0.514 | 0.428 | 
| 62.50 | 0.475 | 0.389 | 
| 31.25 | 0.208 | 0.122 | 
| 0.00 | 0.086 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 90-105% | 97% | 
| EDTA plasma(n=5) | 80-92% | 86% | 
| Heparin plasma(n=5) | 88-102% | 95% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 81-94% | 96-105% | 92-103% | 82-96% | 
| EDTA plasma(n=5) | 89-101% | 87-98% | 92-104% | 82-96% | 
| Heparin plasma(n=5) | 93-102% | 80-93% | 91-100% | 88-95% |