| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 2.168 | 2.084 |
| 100.00 | 1.622 | 1.538 |
| 50.00 | 1.148 | 1.064 |
| 25.00 | 0.899 | 0.815 |
| 12.50 | 0.544 | 0.460 |
| 6.25 | 0.369 | 0.285 |
| 3.13 | 0.234 | 0.150 |
| 0.00 | 0.084 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 78-93% | 85% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-104% | 93-102% | 79-94% | 87-98% |
| EDTA plasma(n=5) | 86-97% | 93-102% | 87-98% | 83-97% |
| Heparin plasma(n=5) | 89-97% | 85-97% | 87-96% | 85-96% |