| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 1.898 | 1.817 |
| 250.00 | 1.407 | 1.326 |
| 125.00 | 0.824 | 0.743 |
| 62.50 | 0.671 | 0.590 |
| 31.25 | 0.386 | 0.305 |
| 15.63 | 0.252 | 0.171 |
| 7.82 | 0.262 | 0.181 |
| 0.00 | 0.081 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 78-92% | 85% |
| Heparin plasma(n=5) | 92-107% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-103% | 83-94% | 92-105% | 89-97% |
| EDTA plasma(n=5) | 88-102% | 81-96% | 86-93% | 82-95% |
| Heparin plasma(n=5) | 90-99% | 95-106% | 87-101% | 82-97% |