| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.984 | 1.881 |
| 500.00 | 1.462 | 1.359 |
| 250.00 | 0.953 | 0.850 |
| 125.00 | 0.606 | 0.503 |
| 62.50 | 0.180 | 0.077 |
| 31.25 | 0.186 | 0.083 |
| 15.63 | 0.213 | 0.110 |
| 0.00 | 0.103 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-107% | 98% |
| EDTA plasma(n=5) | 81-126% | 104% |
| Heparin plasma(n=5) | 87-108% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-105% | 85-94% | 81-105% | 80-101% |
| EDTA plasma(n=5) | 82-90% | 94-114% | 92-118% | 94-108% |
| Heparin plasma(n=5) | 85-104% | 85-97% | 93-113% | 88-105% |