| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.129 | 2.055 |
| 5.00 | 1.498 | 1.424 |
| 2.50 | 1.092 | 1.018 |
| 1.25 | 0.852 | 0.778 |
| 0.63 | 0.558 | 0.484 |
| 0.31 | 0.342 | 0.268 |
| 0.16 | 0.227 | 0.153 |
| 0.00 | 0.074 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-112% | 103% |
| EDTA plasma(n=5) | 92-118% | 105% |
| Heparin plasma(n=5) | 83-98% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-100% | 86-99% | 95-113% | 87-100% |
| EDTA plasma(n=5) | 92-107% | 91-97% | 81-90% | 92-104% |
| Heparin plasma(n=5) | 89-98% | 95-105% | 86-95% | 85-92% |
