| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.401 | 2.305 | 
| 5.00 | 1.563 | 1.467 | 
| 2.50 | 1.108 | 1.012 | 
| 1.25 | 0.835 | 0.739 | 
| 0.63 | 0.531 | 0.435 | 
| 0.32 | 0.379 | 0.283 | 
| 0.16 | 0.214 | 0.118 | 
| 0.00 | 0.096 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 88-106% | 97% | 
| EDTA plasma(n=5) | 81-95% | 88% | 
| Heparin plasma(n=5) | 78-92% | 85% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-92% | 79-96% | 89-100% | 85-96% | 
| EDTA plasma(n=5) | 95-104% | 82-94% | 86-96% | 81-97% | 
| Heparin plasma(n=5) | 92-106% | 92-101% | 80-95% | 87-96% |