| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.266 | 2.181 | 
| 500.00 | 1.679 | 1.594 | 
| 250.00 | 1.243 | 1.158 | 
| 125.00 | 0.864 | 0.779 | 
| 62.50 | 0.595 | 0.510 | 
| 31.25 | 0.368 | 0.283 | 
| 15.63 | 0.225 | 0.140 | 
| 0.00 | 0.085 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 84-102% | 93% | 
| EDTA plasma(n=5) | 87-95% | 90% | 
| Heparin plasma(n=5) | 80-95% | 87% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-94% | 87-101% | 89-99% | 85-95% | 
| EDTA plasma(n=5) | 83-103% | 82-93% | 86-95% | 81-96% | 
| Heparin plasma(n=5) | 92-105% | 92-101% | 80-94% | 86-101% |