| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.043 | 1.978 | 
| 500.00 | 1.731 | 1.666 | 
| 250.00 | 1.087 | 1.022 | 
| 125.00 | 0.735 | 0.670 | 
| 62.50 | 0.518 | 0.453 | 
| 31.25 | 0.363 | 0.298 | 
| 15.63 | 0.100 | 0.035 | 
| 0.00 | 0.065 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 82-95% | 88% | 
| EDTA plasma(n=5) | 85-97% | 91% | 
| Heparin plasma(n=5) | 92-105% | 98% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 88-99% | 81-93% | 89-97% | 85-94% | 
| EDTA plasma(n=5) | 95-103% | 79-97% | 93-102% | 87-103% | 
| Heparin plasma(n=5) | 85-94% | 83-96% | 87-98% | 87-100% |