Products

Mouse I-PTH(IntactParathormone) Microsample ELISA Kit
One-click to copy product information
ELK2564MS
Size: Price:
48 T $320.00
96 T $458.00
96 T*5 $1947.00

Overview

Product name: Mouse I-PTH(IntactParathormone) Microsample ELISA Kit
Reactivity: Mouse
Alternative Names: iPTH; Intact Parathyroid Hormone; Parathormone; Parathyrin; PTH; Parathyroid Hormone
Assay Type: Sandwich
Sensitivity: 3.96 pg/mL
Standard: 800 pg/mL
Detection Range: 12.5-800 pg/mL
Sample Type: serum, plasma and other biological fluids
Assay Length: 3.5h
Research Area: Metabolic pathway;Endocrinology;Hormone metabolism;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse I-PTH. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse I-PTH. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse I-PTH, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse I-PTH in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
800.00 2.155 2.067
400.00 1.543 1.455
200.00 1.051 0.963
100.00 0.894 0.806
50.00 0.479 0.391
25.00 0.389 0.301
12.50 0.164 0.076
0.00 0.088 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant I-PTH and the recovery rates were calculated by comparing the measured value to the expected amount of I-PTH in samples.
Matrix Recovery range Average
serum(n=5) 86-92% 89%
EDTA plasma(n=5) 92-113% 103%
Heparin plasma(n=5) 92-119% 106%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of I-PTH and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 93-104% 81-104% 87-110% 89-106%
EDTA plasma(n=5) 91-113% 85-92% 85-113% 86-94%
Heparin plasma(n=5) 90-105% 85-106% 88-93% 87-107%