| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.188 | 2.093 | 
| 500.00 | 1.700 | 1.605 | 
| 250.00 | 1.148 | 1.053 | 
| 125.00 | 0.869 | 0.774 | 
| 62.50 | 0.535 | 0.440 | 
| 31.25 | 0.364 | 0.269 | 
| 15.63 | 0.188 | 0.093 | 
| 0.00 | 0.095 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 86-101% | 94% | 
| EDTA plasma(n=5) | 85-91% | 88% | 
| Heparin plasma(n=5) | 83-93% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-109% | 81-87% | 83-116% | 88-118% | 
| EDTA plasma(n=5) | 84-116% | 84-116% | 88-107% | 92-112% | 
| Heparin plasma(n=5) | 91-116% | 91-117% | 84-117% | 85-98% | 