| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.075 | 1.989 | 
| 5.00 | 1.502 | 1.416 | 
| 2.50 | 1.067 | 0.981 | 
| 1.25 | 0.872 | 0.786 | 
| 0.63 | 0.593 | 0.507 | 
| 0.32 | 0.315 | 0.229 | 
| 0.16 | 0.252 | 0.166 | 
| 0.00 | 0.086 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 87-99% | 93% | 
| EDTA plasma(n=5) | 85-97% | 91% | 
| Heparin plasma(n=5) | 90-103% | 96% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 82-94% | 95-102% | 87-103% | 88-95% | 
| EDTA plasma(n=5) | 98-107% | 90-99% | 88-97% | 85-97% | 
| Heparin plasma(n=5) | 81-94% | 87-105% | 79-95% | 96-105% |