| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 100.00 | 2.058 | 1.965 | 
| 50.00 | 1.472 | 1.379 | 
| 25.00 | 1.096 | 1.003 | 
| 12.50 | 0.807 | 0.714 | 
| 6.25 | 0.547 | 0.454 | 
| 3.13 | 0.357 | 0.264 | 
| 1.57 | 0.235 | 0.142 | 
| 0.00 | 0.093 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 83-95% | 89% | 
| EDTA plasma(n=5) | 87-99% | 93% | 
| Heparin plasma(n=5) | 92-105% | 98% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 95-106% | 88-103% | 81-98% | 93-102% | 
| EDTA plasma(n=5) | 90-99% | 95-106% | 87-101% | 82-97% | 
| Heparin plasma(n=5) | 88-103% | 83-94% | 92-105% | 89-97% |