| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.172 | 2.091 |
| 2500.00 | 1.562 | 1.481 |
| 1250.00 | 1.225 | 1.144 |
| 625.00 | 0.937 | 0.856 |
| 312.50 | 0.516 | 0.435 |
| 156.25 | 0.369 | 0.288 |
| 78.13 | 0.215 | 0.134 |
| 0.00 | 0.081 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 86-99% | 92% |
| EDTA plasma(n=5) | 79-93% | 86% |
| Heparin plasma(n=5) | 90-105% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-92% | 83-96% | 93-102% | 87-98% |
| EDTA plasma(n=5) | 92-103% | 93-105% | 88-101% | 97-106% |
| Heparin plasma(n=5) | 85-94% | 90-99% | 95-102% | 82-90% |