| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.061 | 1.967 |
| 500.00 | 1.527 | 1.433 |
| 250.00 | 1.120 | 1.026 |
| 125.00 | 0.762 | 0.668 |
| 62.50 | 0.561 | 0.467 |
| 31.25 | 0.340 | 0.246 |
| 15.63 | 0.170 | 0.076 |
| 0.00 | 0.094 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-110% | 103% |
| EDTA plasma(n=5) | 86-128% | 107% |
| Heparin plasma(n=5) | 88-99% | 94% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-114% | 81-120% | 86-119% | 86-108% |
| EDTA plasma(n=5) | 89-102% | 93-107% | 90-118% | 93-111% |
| Heparin plasma(n=5) | 91-97% | 80-115% | 84-109% | 95-100% |