| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.123 | 2.030 |
| 500.00 | 1.659 | 1.566 |
| 250.00 | 1.157 | 1.064 |
| 125.00 | 0.899 | 0.806 |
| 62.50 | 0.567 | 0.474 |
| 31.25 | 0.386 | 0.293 |
| 15.63 | 0.225 | 0.132 |
| 0.00 | 0.093 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-105% | 98% |
| EDTA plasma(n=5) | 82-94% | 88% |
| Heparin plasma(n=5) | 83-99% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-99% | 83-89% | 88-104% | 97-103% |
| EDTA plasma(n=5) | 85-95% | 79-96% | 89-101% | 85-97% |
| Heparin plasma(n=5) | 86-96% | 93-104% | 89-97% | 85-94% |