| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.268 | 2.181 |
| 1000.00 | 1.647 | 1.560 |
| 500.00 | 1.069 | 0.982 |
| 250.00 | 0.863 | 0.776 |
| 125.00 | 0.518 | 0.431 |
| 62.50 | 0.353 | 0.266 |
| 31.25 | 0.218 | 0.131 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-99% | 91% |
| EDTA plasma(n=5) | 79-91% | 85% |
| Heparin plasma(n=5) | 81-97% | 89% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-97% | 88-95% | 92-101% | 87-98% |
| EDTA plasma(n=5) | 92-101% | 82-98% | 89-97% | 85-93% |
| Heparin plasma(n=5) | 86-101% | 86-101% | 87-96% | 92-105% |