| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 6000.00 | 2.248 | 2.144 |
| 3000.00 | 1.589 | 1.485 |
| 1500.00 | 1.122 | 1.018 |
| 750.00 | 0.783 | 0.679 |
| 375.00 | 0.509 | 0.405 |
| 187.50 | 0.360 | 0.256 |
| 93.75 | 0.259 | 0.155 |
| 0.00 | 0.104 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-111% | 98% |
| EDTA plasma(n=5) | 83-109% | 96% |
| Heparin plasma(n=5) | 87-96% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 90-107% | 86-118% | 84-116% | 93-119% |
| EDTA plasma(n=5) | 81-92% | 83-111% | 90-98% | 84-90% |
| Heparin plasma(n=5) | 89-114% | 95-113% | 92-120% | 81-86% |