| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 500.00 | 2.121 | 2.036 | 
| 250.00 | 1.694 | 1.609 | 
| 125.00 | 1.412 | 1.327 | 
| 62.50 | 0.780 | 0.695 | 
| 31.25 | 0.509 | 0.424 | 
| 15.63 | 0.305 | 0.220 | 
| 7.82 | 0.258 | 0.173 | 
| 0.00 | 0.085 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 82-95% | 88% | 
| EDTA plasma(n=5) | 86-96% | 91% | 
| Heparin plasma(n=5) | 89-107% | 98% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 86-93% | 91-101% | 87-98% | 95-104% | 
| EDTA plasma(n=5) | 89-99% | 83-99% | 98-105% | 96-101% | 
| Heparin plasma(n=5) | 85-100% | 84-99% | 85-97% | 88-102% |