Product name: Human ET-1(Endothelin1) Microsample ELISA Kit
Reactivity: Human
Alternative Names: ET1; PPET1; Preproendothelin-1; Big endothelin-1; EDN1
Assay Type: Sandwich
Sensitivity: 2.69 pg/mL
Standard: 500 pg/mL
Detection Range: 7.82-500 pg/mL
Sample Type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 3.5h
Research Area: Endocrinology;Cardiovascular biology;
Uniprot ID: P05305
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human ET-1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human ET-1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human ET-1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human ET-1 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
500.00 2.057 1.951
250.00 1.474 1.368
125.00 1.004 0.898
62.50 0.765 0.659
31.25 0.517 0.411
15.63 0.335 0.229
7.81 0.209 0.103
0.00 0.106 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant ET-1 and the recovery rates were calculated by comparing the measured value to the expected amount of ET-1 in samples.
Matrix Recovery range Average
serum(n=5) 82-112% 97%
EDTA plasma(n=5) 88-116% 102%
Heparin plasma(n=5) 84-109% 97%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of ET-1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 95-118% 90-118% 87-94% 91-97%
EDTA plasma(n=5) 93-102% 82-97% 94-113% 94-118%
Heparin plasma(n=5) 86-113% 82-115% 93-100% 92-116%

Product Components