| 48 T | $412.00 | 
| 96 T | $588.00 | 
| 96 T*5 | $2499.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 1.974 | 1.887 | 
| 500.00 | 1.579 | 1.492 | 
| 250.00 | 1.215 | 1.128 | 
| 125.00 | 0.661 | 0.574 | 
| 62.50 | 0.593 | 0.506 | 
| 31.25 | 0.357 | 0.270 | 
| 15.63 | 0.190 | 0.103 | 
| 0.00 | 0.087 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 78-92% | 85% | 
| EDTA plasma(n=5) | 80-95% | 87% | 
| Heparin plasma(n=5) | 78-96% | 87% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 88-101% | 85-93% | 89-97% | 85-94% | 
| EDTA plasma(n=5) | 87-96% | 92-105% | 89-102% | 92-101% | 
| Heparin plasma(n=5) | 88-102% | 87-102% | 85-96% | 87-101% |