| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.058 | 1.955 |
| 1000.00 | 1.585 | 1.482 |
| 500.00 | 1.111 | 1.008 |
| 250.00 | 0.813 | 0.710 |
| 125.00 | 0.591 | 0.488 |
| 62.50 | 0.364 | 0.261 |
| 31.25 | 0.261 | 0.158 |
| 0.00 | 0.103 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 86-99% | 92% |
| EDTA plasma(n=5) | 89-103% | 96% |
| Heparin plasma(n=5) | 83-97% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-104% | 87-96% | 87-99% | 96-107% |
| EDTA plasma(n=5) | 93-105% | 79-86% | 87-103% | 85-97% |
| Heparin plasma(n=5) | 86-105% | 98-107% | 90-99% | 88-97% |