| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 3000.00 | 2.058 | 1.971 |
| 1500.00 | 1.672 | 1.585 |
| 750.00 | 1.243 | 1.156 |
| 375.00 | 0.963 | 0.876 |
| 187.50 | 0.547 | 0.460 |
| 93.75 | 0.357 | 0.270 |
| 46.88 | 0.190 | 0.103 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-96% | 87% |
| EDTA plasma(n=5) | 86-99% | 92% |
| Heparin plasma(n=5) | 89-103% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-97% | 82-96% | 86-98% | 93-101% |
| EDTA plasma(n=5) | 87-96% | 86-95% | 92-103% | 81-90% |
| Heparin plasma(n=5) | 82-96% | 85-98% | 92-103% | 81-95% |