Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (mIU/mL) | OD | Corrected OD |
---|---|---|
30.00 | 2.096 | 1.992 |
15.00 | 1.708 | 1.604 |
7.50 | 1.211 | 1.107 |
3.75 | 0.828 | 0.724 |
1.88 | 0.563 | 0.459 |
0.94 | 0.368 | 0.264 |
0.47 | 0.219 | 0.115 |
0.00 | 0.104 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 78-90% | 84% |
EDTA plasma(n=5) | 95-107% | 101% |
Heparin plasma(n=5) | 90-105% | 97% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-94% | 87-103% | 87-96% | 82-90% |
EDTA plasma(n=5) | 96-105% | 88-96% | 87-98% | 93-101% |
Heparin plasma(n=5) | 89-97% | 95-104% | 93-102% | 79-94% |