| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (µg/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 1.996 | 1.916 | 
| 5.00 | 1.520 | 1.440 | 
| 2.50 | 1.065 | 0.985 | 
| 1.25 | 0.761 | 0.681 | 
| 0.63 | 0.393 | 0.313 | 
| 0.32 | 0.265 | 0.185 | 
| 0.16 | 0.175 | 0.095 | 
| 0.00 | 0.080 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 78-94% | 86% | 
| EDTA plasma(n=5) | 90-103% | 96% | 
| Heparin plasma(n=5) | 81-96% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 95-102% | 93-101% | 81-89% | 85-99% | 
| EDTA plasma(n=5) | 88-97% | 79-92% | 81-97% | 87-95% | 
| Heparin plasma(n=5) | 85-94% | 96-105% | 87-101% | 82-90% |