 
| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 5000.00 | 2.051 | 1.965 | 
| 2500.00 | 1.594 | 1.508 | 
| 1250.00 | 1.198 | 1.112 | 
| 625.00 | 0.907 | 0.821 | 
| 312.50 | 0.538 | 0.452 | 
| 156.25 | 0.326 | 0.240 | 
| 78.13 | 0.196 | 0.110 | 
| 0.00 | 0.086 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 81-95% | 88% | 
| EDTA plasma(n=5) | 85-97% | 91% | 
| Heparin plasma(n=5) | 91-105% | 98% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 89-105% | 87-98% | 92-100% | 88-104% | 
| EDTA plasma(n=5) | 88-97% | 91-105% | 87-98% | 95-105% | 
| Heparin plasma(n=5) | 84-99% | 83-101% | 86-99% | 93-101% |