| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.117 | 2.025 |
| 500.00 | 1.639 | 1.547 |
| 250.00 | 1.112 | 1.020 |
| 125.00 | 0.865 | 0.773 |
| 62.50 | 0.533 | 0.441 |
| 31.25 | 0.375 | 0.283 |
| 15.63 | 0.189 | 0.097 |
| 0.00 | 0.092 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-102% | 97% |
| EDTA plasma(n=5) | 81-95% | 88% |
| Heparin plasma(n=5) | 95-103% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-100% | 79-91% | 89-103% | 82-95% |
| EDTA plasma(n=5) | 78-92% | 83-96% | 95-107% | 87-102% |
| Heparin plasma(n=5) | 95-104% | 93-102% | 79-94% | 87-98% |