| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.183 | 2.089 |
| 500.00 | 1.598 | 1.504 |
| 250.00 | 1.069 | 0.975 |
| 125.00 | 0.826 | 0.732 |
| 62.50 | 0.556 | 0.462 |
| 31.25 | 0.387 | 0.293 |
| 15.63 | 0.206 | 0.112 |
| 0.00 | 0.094 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-106% | 100% |
| EDTA plasma(n=5) | 87-93% | 90% |
| Heparin plasma(n=5) | 80-90% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-98% | 81-94% | 96-105% | 92-103% |
| EDTA plasma(n=5) | 87-98% | 92-105% | 95-107% | 82-96% |
| Heparin plasma(n=5) | 81-94% | 93-107% | 93-102% | 91-105% |