| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.106 | 2.013 | 
| 500.00 | 1.527 | 1.434 | 
| 250.00 | 1.175 | 1.082 | 
| 125.00 | 0.784 | 0.691 | 
| 62.50 | 0.529 | 0.436 | 
| 31.25 | 0.369 | 0.276 | 
| 15.63 | 0.199 | 0.106 | 
| 0.00 | 0.093 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 80-92% | 86% | 
| EDTA plasma(n=5) | 88-102% | 95% | 
| Heparin plasma(n=5) | 90-105% | 97% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 93-102% | 90-99% | 95-106% | 87-101% | 
| EDTA plasma(n=5) | 88-101% | 93-106% | 85-95% | 92-104% | 
| Heparin plasma(n=5) | 87-99% | 85-96% | 87-101% | 89-102% |