| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.040 | 1.951 |
| 2500.00 | 1.496 | 1.407 |
| 1250.00 | 1.043 | 0.954 |
| 625.00 | 0.769 | 0.680 |
| 312.50 | 0.572 | 0.483 |
| 156.25 | 0.381 | 0.292 |
| 78.13 | 0.168 | 0.079 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-105% | 97% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 78-90% | 84% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-101% | 84-101% | 95-104% | 88-105% |
| EDTA plasma(n=5) | 86-96% | 91-107% | 87-101% | 95-102% |
| Heparin plasma(n=5) | 92-103% | 83-98% | 88-97% | 96-104% |