| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.975 | 1.888 |
| 5.00 | 1.524 | 1.437 |
| 2.50 | 1.142 | 1.055 |
| 1.25 | 0.734 | 0.647 |
| 0.63 | 0.422 | 0.335 |
| 0.31 | 0.326 | 0.239 |
| 0.16 | 0.194 | 0.107 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-113% | 104% |
| EDTA plasma(n=5) | 89-105% | 97% |
| Heparin plasma(n=5) | 94-99% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-101% | 84-117% | 88-120% | 83-112% |
| EDTA plasma(n=5) | 93-98% | 86-109% | 82-108% | 92-107% |
| Heparin plasma(n=5) | 83-97% | 92-104% | 81-97% | 95-116% |