 
| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 500.00 | 1.974 | 1.878 | 
| 250.00 | 1.579 | 1.483 | 
| 125.00 | 1.133 | 1.037 | 
| 62.50 | 0.877 | 0.781 | 
| 31.25 | 0.509 | 0.413 | 
| 15.63 | 0.357 | 0.261 | 
| 7.82 | 0.221 | 0.125 | 
| 0.00 | 0.096 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 88-103% | 95% | 
| EDTA plasma(n=5) | 92-106% | 99% | 
| Heparin plasma(n=5) | 77-95% | 86% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 90-99% | 83-97% | 81-103% | 89-101% | 
| EDTA plasma(n=5) | 85-97% | 87-96% | 82-95% | 92-105% | 
| Heparin plasma(n=5) | 85-97% | 90-99% | 82-101% | 93-105% |