| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 2000.00 | 1.952 | 1.864 | 
| 1000.00 | 1.587 | 1.499 | 
| 500.00 | 1.197 | 1.109 | 
| 250.00 | 0.921 | 0.833 | 
| 125.00 | 0.597 | 0.509 | 
| 62.50 | 0.273 | 0.185 | 
| 31.25 | 0.226 | 0.138 | 
| 0.00 | 0.088 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 87-99% | 94% | 
| EDTA plasma(n=5) | 84-95% | 88% | 
| Heparin plasma(n=5) | 82-97% | 92% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 91-99% | 83-89% | 88-104% | 97-103% | 
| EDTA plasma(n=5) | 95-104% | 93-102% | 79-94% | 87-98% | 
| Heparin plasma(n=5) | 86-93% | 91-105% | 87-98% | 89-94% |