| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.079 | 1.975 |
| 500.00 | 1.636 | 1.532 |
| 250.00 | 1.243 | 1.139 |
| 125.00 | 0.682 | 0.578 |
| 62.50 | 0.593 | 0.489 |
| 31.25 | 0.319 | 0.215 |
| 15.63 | 0.228 | 0.124 |
| 0.00 | 0.104 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 87-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-97% | 97-105% | 81-93% | 89-102% |
| EDTA plasma(n=5) | 87-98% | 92-105% | 83-92% | 89-98% |
| Heparin plasma(n=5) | 81-94% | 96-105% | 92-103% | 82-96% |