| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 500.00 | 2.185 | 2.097 | 
| 250.00 | 1.687 | 1.599 | 
| 125.00 | 1.052 | 0.964 | 
| 62.50 | 0.733 | 0.645 | 
| 31.25 | 0.547 | 0.459 | 
| 15.63 | 0.351 | 0.263 | 
| 7.82 | 0.216 | 0.128 | 
| 0.00 | 0.088 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 87-99% | 93% | 
| EDTA plasma(n=5) | 93-107% | 100% | 
| Heparin plasma(n=5) | 87-99% | 93% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 86-97% | 93-102% | 87-98% | 83-97% | 
| EDTA plasma(n=5) | 82-96% | 85-96% | 87-101% | 89-97% | 
| Heparin plasma(n=5) | 85-101% | 84-98% | 85-93% | 86-98% |