| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.067 | 1.968 |
| 250.00 | 1.661 | 1.562 |
| 125.00 | 1.244 | 1.145 |
| 62.50 | 0.909 | 0.810 |
| 31.25 | 0.525 | 0.426 |
| 15.63 | 0.321 | 0.222 |
| 7.81 | 0.173 | 0.074 |
| 0.00 | 0.099 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 91-111% | 101% |
| EDTA plasma(n=5) | 84-105% | 95% |
| Heparin plasma(n=5) | 89-114% | 102% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-119% | 81-95% | 85-111% | 83-94% |
| EDTA plasma(n=5) | 93-110% | 91-100% | 86-105% | 90-101% |
| Heparin plasma(n=5) | 92-104% | 83-100% | 94-113% | 88-105% |