| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.919 | 1.815 |
| 500.00 | 1.628 | 1.524 |
| 250.00 | 1.253 | 1.149 |
| 125.00 | 0.819 | 0.715 |
| 62.50 | 0.541 | 0.437 |
| 31.25 | 0.293 | 0.189 |
| 15.63 | 0.228 | 0.124 |
| 0.00 | 0.104 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 80-93% | 86% |
| Heparin plasma(n=5) | 94-105% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-94% | 87-105% | 79-95% | 96-105% |
| EDTA plasma(n=5) | 82-94% | 95-102% | 87-103% | 88-95% |
| Heparin plasma(n=5) | 98-107% | 90-99% | 88-97% | 85-97% |