| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 2.260 | 2.156 |
| 10.00 | 1.577 | 1.473 |
| 5.00 | 1.067 | 0.963 |
| 2.50 | 0.756 | 0.652 |
| 1.25 | 0.517 | 0.413 |
| 0.63 | 0.329 | 0.225 |
| 0.32 | 0.228 | 0.124 |
| 0.00 | 0.104 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-105% | 97% |
| EDTA plasma(n=5) | 78-90% | 84% |
| Heparin plasma(n=5) | 84-97% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-93% | 89-97% | 85-94% | 87-101% |
| EDTA plasma(n=5) | 87-93% | 85-95% | 89-99% | 85-95% |
| Heparin plasma(n=5) | 89-97% | 85-94% | 79-96% | 85-99% |