| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 4000.00 | 2.035 | 1.936 | 
| 2000.00 | 1.533 | 1.434 | 
| 1000.00 | 1.139 | 1.040 | 
| 500.00 | 0.752 | 0.653 | 
| 250.00 | 0.548 | 0.449 | 
| 125.00 | 0.363 | 0.264 | 
| 62.50 | 0.196 | 0.097 | 
| 0.00 | 0.099 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 92-106% | 100% | 
| EDTA plasma(n=5) | 87-99% | 93% | 
| Heparin plasma(n=5) | 80-95% | 87% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 78-94% | 87-96% | 85-93% | 88-103% | 
| EDTA plasma(n=5) | 85-92% | 90-96% | 87-1028% | 85-97% | 
| Heparin plasma(n=5) | 88-105% | 86-102% | 97-106% | 89-97% |