| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.154 | 2.089 |
| 2500.00 | 1.593 | 1.528 |
| 1250.00 | 0.997 | 0.932 |
| 625.00 | 0.682 | 0.617 |
| 312.50 | 0.487 | 0.422 |
| 156.25 | 0.239 | 0.174 |
| 78.13 | 0.100 | 0.035 |
| 0.00 | 0.065 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-105% | 97% |
| EDTA plasma(n=5) | 82-94% | 88% |
| Heparin plasma(n=5) | 78-96% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-91% | 87-96% | 95-103% | 85-91% |
| EDTA plasma(n=5) | 85-92% | 90-96% | 95-102% | 87-98% |
| Heparin plasma(n=5) | 88-102% | 93-105% | 91-99% | 83-96% |