| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 2.204 | 2.109 |
| 10.00 | 1.653 | 1.558 |
| 5.00 | 0.943 | 0.848 |
| 2.50 | 0.741 | 0.646 |
| 1.25 | 0.572 | 0.477 |
| 0.63 | 0.389 | 0.294 |
| 0.32 | 0.218 | 0.123 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-106% | 99% |
| EDTA plasma(n=5) | 77-95% | 86% |
| Heparin plasma(n=5) | 88-103% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-94% | 79-96% | 98-105% | 82-93% |
| EDTA plasma(n=5) | 85-92% | 96-104% | 90-98% | 80-92% |
| Heparin plasma(n=5) | 93-101% | 86-92% | 81-93% | 79-95% |